Figure 7-4 Source Exhaust Drain Bottle
Description
Item
Connection to vent
1
Source exhaust drain tubing: 2.5 cm (1.0 inch) inside diameter (i.d.)
2
Roughing pump exhaust hose: 3.2 cm (1.25 inch) i.d.
3
Source exhaust drain bottle. Make sure that the bottle is secured to prevent spillage.
4
Connection to the mass spectrometer: 1.6 cm (0.625 inch) i.d.
5
Roughing pump vacuum inlet hose
6
Note:
Source exhaust hose connections at the drain bottle, mass spectrometer, and the lab vent are secured
with hose clamps.