Description
Label
The items in this list specify how the sample for a given analyte should be grouped
for the calculation of the statistics. The following choices are available:
•
Group by Concentration for Standards
: Standard samples are grouped
by actual concentration.
•
Group by Concentration for QCs
: Quality Control samples are grouped by
actual concentration.
•
Group by Sample Name for Standards
: Replicate Standard samples are
grouped by the Sample Name field.
•
Group by Sample Name for QCs
: Replicate Quality Control samples are
grouped by the Quality Control field.
•
Group by Sample Name for All Samples
: All replicate samples are grouped
by the Sample Name field.
Group by Concentration
for ...
The items in this list specify the actual metric that is used for the calculation of the
statistics. The following choices are available:
•
Calculated Concentration
: The Calculated Concentration field of the Results
Table is used.
•
Area
: The Area field of the Results Table is used.
•
Height
: The Height field of the Results Table is used.
•
Calibration Y-Value
: The regression parameter specified for the analyte is
used. This is either Area or Height for an analyte that does not have a
corresponding internal standard, or Area Ratio or Height Ratio for an analyte that
uses an internal standard.
Calculated Concentration
Analyze Data Using Statistics
Use the statistics pane to view information related to the reproducibility of an analysis. Each row of the table
summarizes information, such as the average and standard deviation, for a group of related peaks from the same
analyte that would ideally be expected to have the same response.
Review the peak integration, the calibration curve, and the sample statistics using an iterative process. The precision
set for the Actual Concentration field in the Results Table is used in the statistics table as well.
Note:
Refer to the laboratory standard operating procedures for information about accepted values for the
statistics, including %CV and Accuracy.
• Open a Results Table and then click
Views > Statistics pane.
System User Guide
SCIEX X500 QTOF System
RUO-IDV-05-2334-E
178 / 234
Operating Instructions—Software