Figure 6-37 Process All Overlays? Dialog
7. If the Process All Overlays? dialog opens, then do one of the following:
• Select
All Overlaid
to generate overlaid XICs for all of the available samples.
• Select
Active Only
to generate XICs only from the currently active sample.
If the
Only show this dialog again if the Shift key is down
check box is selected, then the selected
action is always used unless the user holds the
Shift
key to change the option.
Baseline Subtract Chromatogram
Removes a relatively slowly varying background from a chromatogram.
For each data point in the chromatogram, a window is centered at the corresponding x-value and the points with
minimum intensity within the window to the left and right are found. A straight line is joined between these two
points and the y-value is calculated at the center of the window. This is the baseline that is removed from the data
at that point.
1. Open the Explorer workspace.
2. Click
File > Open Multiple Samples.
SCIEX X500 QTOF System
System User Guide
135 / 234
RUO-IDV-05-2334-E
Operating Instructions—Software