RF627, RF627Smart [Revision 2.1.2] 20.09.2021
Laser Scanners
15
3. Parameterization tabs.
4. Visualization area.
5. Control buttons and notifications.
Area 1
contains the scanner name, scanner model, serial number, firmware version
and measuring ranges. The scanner name can be changed by the user.
Area 2
contains the following indicators:
Group
Icon
Description
Ethernet
If the connection is established, the
Link
field and the connection
speed value will be displayed. The
Required
field displays the
recommended connection speed required for correct operation.
This message appears when there are delays in the network during
data transfer.
If the connection with the scanner is lost (for example, when the
scanner is restarted or the connection is broken), the web page will
be displayed, but the connection status will be changed to
Disconnected
.
Temp
The processor temperature (
CPU, °C
) and the temperature inside the
scanner body (
Internal, °C
).
The processor temperature (
CPU
) and the internal temperature of the
scanner (
Internal
) in °C. This information is used to assess the
operating conditions of the scanner. Do not allow the temperature to
rise to 90°C or more. The indication turns on when the temperature
rises above 90°C, or if the temperature is below -15°C:
Profiles
Displays the following parameters: the synchronization source (Icon),
the current number of profiles per second (PPS) and the current
format of the profile data (Format) sent by the scanner via UDP.
Synchronization sources:
·
Internal
- Synchronization by the internal generator of the
scanner.
·
External
- Synchronization by the external trigger.
·
Soft
- Synchronization by the software request.
View of buttons when hovering over the cursor:
Counters
Displays the value of the profile counter (
Profile
), the value of the
pulse counter of the encoder (
Pulse
), the direction of the encoder
(
Direction
). On the right side is a button to reset the counters to
zero.
Dump
The level of internal memory for recording profiles and the record
button (
Record
). Recording is possible only for calibrated profiles
(
Data format > Profile
), otherwise the record button is not active.
Inputs
The status of the scanner inputs. Waveforms of digital signals at the
inputs. Waveforms are only displayed for enabled inputs.
Area 3
provides access to the scanner settings and includes the following tabs:
Содержание RF627 Series