VNA GUI reference
R&S
®
ZNL/ZNLE
374
User Manual 1178.5966.02 ─ 19
●
Frequency list
If the cable attenuation is specified at several frequency points, the R&S
ZNL/
ZNLE calculates the
Attenuation (f
center
)
by linear interpolation. If
f
center
is below
(above) the specified frequency range, then
Attenuation (f
center
)
is linearly extrapo-
lated from the two lowest (highest) frequency points.
It is recommended to specify the cable attenuation as accurately as possible. The cor-
rection due to the attenuation factor is proportional to the measured distance between
the fault and the reference plane: The larger the distance, the larger the correction.
Add / Delete
Creates/deletes a user-defined cable type.
Remote command:
CALCulate<Chn>:TRANsform:DTFault:DEFine
CALCulate<Chn>:TRANsform:DTFault:DELete
MMEMory:LOAD:CABLe
MMEMory:STORe:CABLe
Import Cable Type
Opens the "Import Cable Type" variant of the
cable type file (
*.rsc
) for import.
Remote command:
10.6.7.3
Fault List dock widget
The "Fault List" dock widget allows you to explore and save all peaks that violate the
active
Access:
[Trace] > "Distance to Fault" > "Fault List..."
Trace softtool