Command reference
R&S
®
ZNL/ZNLE
653
User Manual 1178.5966.02 ─ 19
SDATa
Output as unformatted trace data; see
. Query only.
FSIData
Output or input at "Fixture Simulation Input" data access point.
<Data>
Trace data either in ASCII or block data format, depending on
the current
The column order must match the one used during export.
Example:
Suppose that a TOSM calibration for ports 1 and 2 is active in
channel no. 1.
CALCulate:DATA:CALL:CATalog?
Query the traces available for
CALCulate<Ch>:DATA:CALL?
.
The response is
'S11,S12,S21,S22'
.
CALCulate:DATA:CALL? SDATa
Return the complex response values of all traces. The traces in
the catalog list are read one after another: The response array
contains n (number of points) pairs of real and imaginary values
for S
11
, followed by n pairs of values for S
12
, S
21
, and S
22
.
CALCulate<Ch>:DATA:CALL:CATalog?
Returns all traces which are available for
in channel
no.
<Ch>
. The response is a string parameter with all S-parameter traces in the current
channel or in the active system error correction; see example.
Suffix:
<Ch>
.
Channel number
Example:
See
Usage:
Query only
CALCulate<Ch>:DATA:CHANnel:ALL?
<Format>
Reads the current response values of all traces of the selected channel.
Suffix:
<Ch>
.
Channel number
Query parameters:
<Format>
FDATa | SDATa | MDATa
Output format for the S-parameter data, see
Return values:
<Data>
Trace data either in ASCII or block data format, depending on
the current
Usage:
Query only
VNA command reference