Command reference
R&S
®
ZNL/ZNLE
1005
User Manual 1178.5966.02 ─ 19
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess:SUBModule:ALL?
If the calibration unit selected using
SYSTem:COMMunicate:RDEVice:AKAL:
(ICS), this query returns the addresses of the
connected ICUs. Otherwise it returns the empty string.
Example:
:SYSTem:COMMunicate:RDEVice:AKAL:
ADDRess 'ZN-Z30-02::100010'
Selects an ICS via its Inline Calibration Controller (ICC)
R&S
ZN-Z30 (serial number 100010).
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess:
SUBModule:ALL?
Returns something like:
'[1]ZN-Z33-02::101325,[2]ZN-Z33-02::
101325,[3]ZN-Z33-02::101339'
i.e. there are three ICUs of type R&S
ZN-Z32 variant 02 connec-
ted to the ICC. For correct port assignment during calibration,
the ICU with serial number 101325 is considered port 1, 101325
port 2, and 101339 port 3 of the ICS.
Usage:
Query only
SYSTem:COMMunicate:RDEVice:AKAL:CKIT
<Characterization>
If the calibration unit selected using
SYSTem:COMMunicate:RDEVice:AKAL:
(ICU), this command sets/queries the characteri-
zation to be used for calibrations with this ICU.
Parameters:
<Characterization>
Name of a characterization of the related ICU
Example:
// Select the ICS
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess 'ZN-Z30-02::100010'
// Query for the connected ICUs
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess:SUBModule:ALL?
// '[1]ZN-Z33-02::101325,[2]ZN-Z33-02::101336,[3]ZN-Z33-02::101339'
// Activate ICU[1]
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess 'ZN-Z33-02::101336'
// Query the available characterizations
:SYSTem:COMMunicate:RDEVice:AKAL:CKIT:CATalog?
// 'Factory,afterAwhile'
// Select the non-factory characterization
:SYSTem:COMMunicate:RDEVice:AKAL:CKIT 'afterAwhile'
// Query the active characterization
:SYSTem:COMMunicate:RDEVice:AKAL:CKIT?
// 'ZN-Z33-02::101336=afterAwhile'
Manual operation:
See
VNA command reference