Command reference
R&S
®
ZNL/ZNLE
826
User Manual 1178.5966.02 ─ 19
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:Y:OFFSet
<MagnitudeFactor>[,
<PhaseFactor>[, <RealPart>[, <ImaginaryPart>]]]
Modifies all points of the trace
<WndTr>
by means of an added and/or a multiplied
complex constant. The response values M of the trace are transformed according to:
Suffix:
<Wnd>
.
Number of an existing diagram area (defined by means of
ON
).
<WndTr>
Existing trace number, assigned by means of
WINDow<Wnd>]:TRACe<WndTr>:FEED
Parameters:
<MagnitudeFactor>
Multiplied magnitude factor
Range:
-300 dB to + 300 dB
*RST:
0 dB
Default unit: dB
<PhaseFactor>
Multiplied phase factor, optional for setting command but
returned by query
Range:
-3.4*1038 deg to +3.4*1038 deg
*RST:
0 deg
Default unit: deg
<RealPart>
Real and imaginary part of added complex constant, optional for
setting command but returned by query
Range:
-3.4*1038 to +3.4*1038
*RST:
0
<ImaginaryPart>
Example:
*RST; :DISP:WIND:TRAC:X:OFFS 1MHZ; :DISP:WIND:
TRAC:Y:OFFS 10
Create the default trace and shift it horizontally by 1 MHz, verti-
cally by 10 dB.
DISP:WIND:TRAC:Y:OFFS?
Query all response offset values. The response is
10,0,0,0
.
Manual operation:
See
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:Y[:SCALe]:AUTO
<Activate>[,
<TraceName>]
Displays the entire trace in the diagram area, leaving an appropriate display margin.
The trace can be referenced either by its number
<WndTr>
or by its name
<TraceName>
.
Suffix:
<Wnd>
.
Number of an existing diagram area (defined by means of
ON
). This suffix is
ignored if the optional
<TraceName>
parameter is used.
VNA command reference