VNA GUI reference
R&S
®
ZNL/ZNLE
505
User Manual 1178.5966.02 ─ 19
Controls in the Characterize Cal Unit dialog
Calibration Unit
Displays the connected calibration units. The R&S
ZNL/ZNLE auto-detects all calibra-
tion units which are connected to one of its USB ports. If several cal units are connec-
ted, one of them must be selected for characterization (active cal unit).
Remote command:
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess:ALL?
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess
Authentication
Allows you to set a password to protect the characterization dialog and the
from unauthorized access and operation. "Set Password..." opens a dia-
log to enter the password and activate password protection at the next time the "Set
Password" dialog is opened. Enter an empty string (no password) to deactivate pass-
word protection.
Tip:
A password also blocks a switchover of the active characterization during calibra-
tion; see
Remote command:
[SENSe:]CORRection:COLLect:AUTO:CKIT:PASSword
Characterization Data
Displays all characterizations which are stored on the active cal unit. The "Factory"
characterization is available for all calibration units; it ensures an accurate calibration
for all standard applications. Characterizations stored on an SD card (inserted at the
cal unit) are prefixed with "SD:".
Tip:
Characterizations stored on an SD card (inserted at the cal unit) are prefixed with
"SD:".
Cal softtool