Command reference
R&S
®
ZNL/ZNLE
901
User Manual 1178.5966.02 ─ 19
Example:
CORR:COLL:AUTO:CKIT 'AutoCalChar'
Generate a cal kit file
AutoCalChar.calkit
for the active cali-
bration unit and store it in the calibration unit.
SENSe:CORRection:COLLect:AUTO:CKIT 'SD:test1'
Generate a cal kit file
test1.calkit
for the active calibration
unit and store it in the calibration unit.
Usage:
Setting only
Manual operation:
See
"File name / Comment (Optional)"
[SENSe:]CORRection:COLLect:AUTO:CKIT:PASSword
<Password>
Enters a password to enable a single password-protected action in the automatic cali-
bration or in the characterization wizard. If password protection has been activated
manually in the "Characterize Cal Unit" dialog, the password is required for any single
execution of one the following commands, provided that a cal kit (characterization) file
different from the active characterization is specified:
[SENSe]:CORRection:COLLect:AUTO
[SENSe]:CORRection:COLLect:AUTO:CKIT
[SENSe]:CORRection:COLLect:AUTO:PORTs
[SENSe]:CORRection:COLLect:AUTO:PORTs:TYPE
[SENSe]:CORRection:COLLect:AUTO:TYPE
Tip:
You have to send
[SENSe:]CORRection:COLLect:AUTO:CKIT:PASSword
repeatedly if your command script uses several password-protected commands. You
do not need a password to perform automatic calibrations using the active cal unit
characterization.
Setting parameters:
<Password>
Password (string parameter), as defined in manual control.
Example:
SENSe:CORRection:COLLect:AUTO:CKIT:PASSword
'My_password'
Enter a password
My_password
(assuming that password pro-
tection has been activated manually).
SENSe:CORRection:COLLect:AUTO '', 1, 2
Perform an automatic 2-port calibration at test ports 1 and 2
using the calibration unit's default calibration kit file and auto-
matic port assignment.
SENSe:CORRection:COLLect:AUTO:CKIT:PASSword
'My_password'
Re-enter the password.
SENSe:CORRection:COLLect:AUTO '', 1, 2
Repeat the calibration.
Usage:
Setting only
Manual operation:
See
VNA command reference