VNA GUI reference
R&S
®
ZNL/ZNLE
509
User Manual 1178.5966.02 ─ 19
File name / Comment (Optional)
Selects a filename to reference the characterization data set in the "Characterize Cal
Unit" and "Calibration Unit" dialogs and a comment, to be written into the characteriza-
tion file. A filename is required before you can "Finish" the characterization wizard and
store the data.
Remote command:
[SENSe:]CORRection:COLLect:AUTO:CKIT
10.12.3
Power Cal Settings tab
R&S
ZNL only
Gives access to all functions for power meter and power calibration data handling
(transmission coefficients). Power calibration is a fully menu-guided process.
Efficient power calibration procedure
●
For standard applications, open the "Start Cal" tab and select "Scalar Power Cal" –
"Power Cal..." to perform the necessary calibration sweeps with default power cali-
bration settings. You do not need any of the buttons in the "Power Cal Settings"
tab.
●
Select "Cal Power..." if you use an amplifier between the source port and the DUT.
●
Select "Transm. Coefficients..." if you want to modify the power calibration proce-
dure.
Background information
Refer to
Chapter 9.5.6, "Scalar power calibration"
Cal softtool