Command reference
R&S
®
ZNL/ZNLE
605
User Manual 1178.5966.02 ─ 19
CALibration:DUE:SHUTdown
<State>
If activated, the R&S
ZNL/ZNLE is automatically shut down after self-alignment is com-
pleted. Note that the instrument cannot be restarted via remote control.
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
0
Example:
//Schedule a self-alignment every Sunday at 2 AM after a
warmup period, then shut down instrument.
CAL:DUE:WARM
CAL:DUE:SCH ON
CAL:DUE:DAYS SUND
CAL:DUE:TIME '2:00'
CAL:DUE:SHUT
CALibration:DUE:TIME
<Time>
Defines the time at which a self-alignment is scheduled for the days specified by
ON
.
Parameters:
<Time>
string with format 'hh:mm' (24 hours)
Example:
//Schedule a self-alignment every Sunday at 2 AM after a
warmup period, then shut down instrument.
CAL:DUE:WARM
CAL:DUE:SCH ON
CAL:DUE:DAYS SUND
CAL:DUE:TIME '2:00'
CAL:DUE:SHUT
CALibration:DUE:WARMup
<State>
If enabled, self-alignment is started automatically after the warmup operation has com-
pleted.
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
0
Common instrument commands