Getting Started
R&S
®
ZNL/ZNLE
70
User Manual 1178.5966.02 ─ 19
6. To obtain a more conclusive peak list that does not contain noise peaks, for exam-
ple, define a threshold that is higher than the noise floor:
a) Press the [MKR] key on the front panel.
b) Tap the "Marker Config" softkey in the "Marker" menu.
c) Tap the "Search" tab in the "Marker" dialog box.
d) In the "Threshold" field, enter
-68
dBm
.
e) Tap the "State" box for "Threshold" to activate its use.
Only peaks that are larger than -68
dBm will be included in the peak list.
The marker peak list displays the determined peaks that are above the defined
threshold.
Figure 4-15: Marker Peak List
4.3.7
Performing sequential measurements
Although only one measurement can be performed at any one time, the measurements
configured in the active channel setups can be performed sequentially, that means:
one after the other, automatically, either once or continuously.
1. Tap the "Sequencer" icon in the toolbar.
2. Toggle the "Sequencer" softkey in the "Sequencer" menu to "On".
A continuous sequence is started, i.e. each channel setup measurement is per-
formed one after the other until the Sequencer is stopped.
Trying out the instrument