VNA GUI reference
R&S
®
ZNL/ZNLE
490
User Manual 1178.5966.02 ─ 19
Apply
Is enabled when a new set of power calibration data has been acquired. The button
applies all available power calibrations to the active channel, aborts the verification
sweeps, and closes the port overview section.
The power calibration state is indicated in the trace list, see
on page 275. Use the functions in the
on page 520 to activate, deactivate, or store power calibrations.
Remote command:
SOURce<Ch>:POWer:CORRection[:ACQuire]
[SENSe<Ch>:]CORRection:POWer<PhyPt>:ACQuire
SOURce<Ch>:POWer:CORRection:DATA
[SENSe<Chn>:]CORRection:PSTate?
Power Cal dialog – Ref. Receiver
In Ref. Receiver mode, the "Power Cal" dialog guides you through a reference receiver
calibration.
The selected source port is displayed with its cal power settings (see
ter 10.12.3.2, "Cal Power Config dialog"
on page 512). A circuit diagram visualizes the
connectivity. A power meter must be connected to the calibrated source port.
Access:
Background information
Refer to
Chapter 9.5.6.1, "Source power calibration"
Cal softtool