VNA GUI reference
R&S
®
ZNL/ZNLE
363
User Manual 1178.5966.02 ─ 19
Resolution Enh.
Broadens the frequency range that the analyzer considers for the time domain trans-
form by a linear factor. A factor of 1 means that the original sweep range and the mea-
sured sweep points are used; no additional assumptions are made. With higher resolu-
tion enhancement factors, the measurement data is extrapolated using a linear predic-
tion method. As a result, the resolution in time domain can be improved.
The ideal resolution enhancement factor depends on the properties of the DUT. For
distance to fault measurements on cables, set it to 1.
Remote command:
CALCulate<Chn>:TRANsform:TIME:RESolution:EFACtor
10.6.5.2
Low Pass Settings dialog
The "Low Pass Settings" dialog defines the harmonic grid for low pass time domain
transforms.
Access:
[Trace] > "Time Domain" > "Low Pass Settings..."
Background information
Refer to
Chapter 9.7.1.4, "Harmonic grid"
Is the Current Grid Harmonic?
The area at the top of the "Low Pass Settings" dialog indicates whether the current fre-
quency grid is harmonic.
Remote command:
Trace softtool