Command reference
R&S
®
ZNL/ZNLE
697
User Manual 1178.5966.02 ─ 19
Parameters:
<Boolean>
ON - show the bandfilter search results. If no bandfilter search
has been initiated before (
BFILter
), nothing is displayed.
OFF - hide the bandfilter search results.
*RST:
OFF
Example:
See
CALCulate<Chn>:MARKer<Mk>:BWIDth
Manual operation:
See
CALCulate<Chn>:MARKer:SEARch:BFILter:RESult[:STATe]:AREA
<HorizontalPos>, <VerticalPos>
Moves the bandfilter search info field for the active trace
<Chn>
to one of nine prede-
fined positions in the active diagram.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<HorizontalPos>
LEFT | MID | RIGHt
Horizontal position
<VerticalPos>
TOP | MID | BOTTom
Vertical position
Example:
See
CALCulate<Chn>:MARKer<Mk>:BWIDth
Manual operation:
See
CALCulate<Chn>:MARKer<Mk>:SEARch:FORMat
<SearchFormat>
CALCulate<Chn>:MARKer<Mk>:REFerence:SEARch:FORMat
<SearchFormat>
Defines the format in which you want to specify the target value
Each marker may have a different target format. The table below gives an overview on
how a complex target value z = x + jy is converted.
Target Format
Description
Formula
MLINear
Magnitude of z, unconverted.
|z| = sqrt ( x
2
+ y
2
)
MLOGarithmic
Magnitude of z in dB
Mag(z) = 20 log|z| dB
PHASe
Phase of z
φ (z) = arctan (y/x)
UPHase
Unwrapped phase of z comprising
the complete number of 360°
phase rotations
Ф(z) = φ (z) + 2k·360°
REAL
Real part of z
Re(z) = x
IMAGinary
Imaginary part of z
Im(z) = y
VNA command reference