VNA GUI reference
R&S
®
ZNL/ZNLE
404
User Manual 1178.5966.02 ─ 19
Background information
Refer to the following sections:
●
Chapter 9.3.5.1, "Wave quantities"
●
The notation for ratios follows the usual scheme of the vector network analyzer:
●
The a-waves are the outgoing/transmitted waves at the analyzer's test ports.
●
The b-waves are the incoming/measured waves.
●
The source port for the stimulus signal must be specified in addition.
●
The port number range covers all test ports of the analyzer.
Numerator
Selects the type (left pull-down list) and the port number assignment (right pull-down
list) of the wave that forms the numerator of the
Remote command:
CALCulate<Ch>:PARameter:MEASure
"<Trace_Name>", "B2/A1" | ...
CALCulate<Ch>:PARameter:SDEFine
"<Trace_Name>", "B2/A1" | ...
Denominator
Selects the type (left pull-down list) and the port number assignment (right pull-down
list) of the wave that forms the denominator of the
Remote command:
CALCulate<Ch>:PARameter:MEASure
"<Trace_Name>", "B2/A1" | ...
CALCulate<Ch>:PARameter:SDEFine
"<Trace_Name>", "B2/A1" | ...
Port
Selects the source port for the stimulus signal ("Port 1" or "Port 2").
Remote command:
CALCulate<Ch>:PARameter:MEASure
"<Trace_Name>", "<Ratio>"
CALCulate<Ch>:PARameter:SDEFine
"<Trace_Name>", "<Ratio>"
Meas softtool