Command reference
R&S
®
ZNL/ZNLE
747
User Manual 1178.5966.02 ─ 19
Note:
If the x-axis is scaled in distance units (
CALCulate<Chn>:TRANsform:TIME:
DISTance
), then the stop value is entered in m; the range and default value
changes accordingly.
CALCulate<Chn>:TRANsform:TIME[:TYPE]
<TransformType>
Selects the time domain transformation type.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<TransformType>
BPASs | LPASs
BPASs - band pass impulse (only impulse response; a step
response
CALCulate<Chn>:TRANsform:TIME:STIMulus
STEP
is automatically changed to impulse response)
LPASs - low pass (impulse or step response, depending on
CALCulate<Chn>:TRANsform:TIME:STIMulus
*RST:
BPASs
Example:
*RST; :CALC:TRAN:TIME:STAT ON
Reset the instrument, activating a frequency sweep, and enable
the time domain transformation for the default trace.
CALC:TRAN:TIME LPAS; TIME:STIM STEP
Select a low pass step transformation.
CALC:TRAN:TIME:LPAS KFST
Calculate a harmonic grid, keeping the stop frequency and the
number of points.
Manual operation:
See
CALCulate<Chn>:TRANsform:TIME:WINDow
<WindowType>
Selects the window type for filtering the data in the frequency domain prior to the time
domain transformation.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<WindowType>
RECT | HAMMing | HANNing | BOHMan | DCHebyshev
RECT - no profiling (rectangle)
HANN - normal profile (Hann)
HAMMing - low first sidelobe (Hamming)
BOHMan - steep falloff (Bohman)
DCHebyshev - arbitrary sidelobes (Dolph-Chebychev)
*RST:
HANN
Example:
See
CALCulate<Chn>:TRANsform:TIME:DCHebyshev
Manual operation:
See
VNA command reference