Command reference
R&S
®
ZNL/ZNLE
883
User Manual 1178.5966.02 ─ 19
[SENSe<Ch>:]BANDwidth[:RESolution]:SELect
<Selectivity>
[SENSe<Ch>:]BWIDth[:RESolution]:SELect
<Selectivity>
Defines the selectivity of the IF filter for an unsegmented sweep. The value is also
used for all segments of a segmented sweep, provided that separate selectivity setting
is disabled
(
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:SELect:CONTrol
OFF
).
Note:
For the R&S
ZNLE always NORMal selectivity filters are used. Trying to set the
selectivity will result in an execution error.
Suffix:
<Ch>
.
Channel number
Parameters:
<Selectivity>
NORMal | MEDium | HIGH
NORMal – IF filter with normal selectivity and short settling time.
MEDium – IF filter with steeper edges and longer settling time.
HIGH – IF filter with highest selectivity but longest settling time.
*RST:
NORMal
Example:
See
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:
Manual operation:
See
[SENSe:]CORRection:CKIT...
The
[SENSe:]CORRection:CKIT...
commands deal with calibration kits and cal kit
data. The calibration kits are distinguished by their names (<CalkitName>), the optional
labels (<label>) can be used to carry information about the calibration standard.
In order to handle several identical calibration kits with different serial numbers use the
commands of
"[SENSe:]CORRection:CKIT... with Labels"
[SENSe:]CORRection:CKIT:<StandardType>
.................................................................... 884
[SENSe:]CORRection:CKIT:CATalog?
.............................................................................. 887
[SENSe:]CORRection:CKIT:DELete
................................................................................. 887
................................................................................ 888
................................................................................... 888
[SENSe:]CORRection:CKIT:SELect
................................................................................. 889
[SENSe:]CORRection:CKIT:<ConnType>:SELect
.............................................................. 889
[SENSe:]CORRection:CKIT:STANdard:CATalog?
.............................................................. 890
[SENSe:]CORRection:CKIT:STANdard:DATA?
.................................................................. 890
VNA command reference