Command reference
R&S
®
ZNL/ZNLE
888
User Manual 1178.5966.02 ─ 19
Manual operation:
See
"Add / Copy / Delete / Standards..."
[SENSe:]CORRection:CKIT:DMODe
<ConnectorType>, <CalKitName>, <KitLabel>,
<Mode>
Sets/gets the delay mode for the related cal kit (identified by connector type, name and
label), i.e allows to toggle between ZVR compatible and Keysight modelling (see
on page 503). Subsequent standard definitions interpret the specified
<DelayParam> accordingly.
In "set direction", if a cal kit with the given connector type, name and label is not availa-
ble on the analyzer, it is created automatically.
Cal kits that are not created/modified using
[SENSe:]CORRection:CKIT:DMODe
use
the ZVR compatible modelling.
Parameters:
<ConnectorType>
Name of the connector type. Use
]CORRection:CONNection:CATalog?
names.
<CalKitName>
String parameter containing the name of the calibration kit. See
"Cal kit naming conventions " under
<KitLabel>
String parameter containing the label of the calibration kit, usu-
ally the serial number.
Setting parameters:
<Mode>
DELay | ELENgth
DELay – Keysight modelling
ELENgth – ZVR compatible modelling
Manual operation:
See
"Add / Copy / Delete / Standards..."
[SENSe:]CORRection:CKIT:LABel
<CalKitName>[, <KitLabel>]
Assigns a label to an imported or user-defined calibration kit.
Parameters:
<CalKitName>
String parameter containing an imported or user-defined calibra-
tion kit available on the analyzer.
<KitLabel>
String parameter containing the calibration kit label.
VNA command reference