Command reference
R&S
®
ZNL/ZNLE
889
User Manual 1178.5966.02 ─ 19
Example:
CORR:CKIT:FOP 'N 50 Ohm','New Kit 1', '',
0,4000000000,0,0,50,99,-2.3,0.22,0,0,0,0,0
Create a new cal kit "New Kit 1" and assign an open (f) standard
for the N 50 Ω connector type with specific properties.
CORR:CKIT:LAB 'New Kit 1', 'Test kit created
today'
Label the previously created kit.
CORR:CKIT:LAB? 'New Kit 1'
Check the label.
CORR:CKIT:DEL 'New Kit 1'
Delete the kit.
[SENSe:]CORRection:CKIT:SELect
<ConnectorType>[, <CalKitName>]
Selects the calibration kit to be used, specifying its connector type and name
(optional).
Tip:
The command is suitable for connector types with arbitrary, user-defined names.
For standard connector types you can use the command
.
Parameters:
<ConnectorType>
Connector type, e.g. a user-defined connector type (string vari-
able).
<CalKitName>
String parameter containing the name of a calibration kit availa-
ble on the analyzer.
Example:
MMEM:LOAD:CKIT 'C:\Users\Public\
Documents\Rohde-Schwarz\ZNL\
Calibration\Kits\New Kit 1.calkit'
Load the previously created cal kit file
New Kit 1.calkit
from the default cal kit directory.
CORR:CKIT:SEL 'N 50 Ohm', 'New Kit 1'
Assign the imported kit to the N 50 Ω connector type, assuming
that the cal kit name stored in
New Kit 1.calkit
reads
New
Kit 1
.
Manual operation:
See
[SENSe:]CORRection:CKIT:<ConnType>:SELect
<CalKitName>
Selects the calibration kit to be used for a specified connector type
<ConnType>
. The
kit is identified by its name.
Tip:
For connector types with arbitrary, user-defined names you can use the command
[SENSe:]CORRection:CKIT:SELect
.
Cal kit naming conventions
VNA command reference