Getting Started
R&S
®
ZNL/ZNLE
62
User Manual 1178.5966.02 ─ 19
4.3.5
Activating additional channel setups
The R&S
ZNL/ZNLE features multiple channel setups, i.e. you can define several mea-
surement configurations in parallel and then switch between the channel setups auto-
matically to perform the measurements sequentially. We will demonstrate this feature
by activating additional channel setups for a different frequency range, a Spectrum
measurement, and an I/Q analysis (note that the latter two measurements are only
available if the optional Spectrum mode is installed).
To activate additional channel setups
1. Press the [Mode] key on the front panel.
2. On the "New Channel Setup" tab of the "Mode" dialog box, tap the "VNA" button.
3. Change the frequency range for this measurement:
Set the
center frequency
to
500
MHz
and the
span
to
1
GHz
.
4. Press the [Mode] key on the front panel.
5. On the "New Channel Setup" tab of the "Mode" dialog box, tap the "Spectrum" but-
ton.
Trying out the instrument