Command reference
R&S
®
ZNL/ZNLE
847
User Manual 1178.5966.02 ─ 19
<StandardType>
MMTHrough | MFTHrough | FFTHrough | MMLine | MMLine1 |
MMLine2 | MMLine3 | MFLine | MFLine1 | MFLine2 | MFLine3 |
FFLine | FFLine1 | FFLine2 | FFLine3 | MMATten | MFATten |
FFATten | MMSNetwork | MFSNetwork | FFSNetwork | MOPen |
FOPen | MSHort | FSHort | MOSHort | MOSHort1 | MOSHort2 |
MOSHort3 | FOSHort | FOSHort1 | FOSHort2 | FOSHort3 |
MREFlect | FREFlect | MMTCh | FMTCh | MSMatch | FSMatch
Standard types; for a description refer to table
.
<StandardLabel>
String parameter addressing a particular calibration standard by
its label. An empty string means that no label is defined.
<TouchstoneFile>
String parameter to specify name and directory of the Touch-
stone file to be loaded. A
*.s1p
file must be used for one-port
standards, a
*.s2p
file for two-port standards.
If no path is specified the analyzer searches the current direc-
tory, to be queried with
*RST:
-
<FirstPort>
First port number (sufficient for one-port standards). If the port
numbers are omitted, the cal kit data is valid for all ports.
<SecondPort>
Second port number, for two-port standards. If the port numbers
are omitted, the cal kit data is valid for all ports.
Example:
MMEM:LOAD:CKIT:SDAT 'N 50 Ohm','Default
Kit',MOPEN,'Test data','test.s1p',1
Load the file Test.s1p from the current directory in order to
define the properties of an Open (m) standard in the cal kit
named "Default Kit" for the N 50 Ω connector type. Assign the
label "Test data" and specify that the standard data is valid for
port no. 1 only.
Usage:
Setting only
Manual operation:
See
"Read .s1p File... / Read .s2p File... /"
Connector and cal kit naming conventions:
Connector and calibration kit names must be entered as string parameters. The strings
contain the connector and cal kit names used in the
Calibration Kits
dialog; a
Ω
in the name must be replaced by 'Ohm', e.g.:
–
'NewKit1'
denotes the user-defined calibration kit "NewKit1".
–
'N 50 Ohm Ideal Kit'
denotes the "N 50
Ω
Ideal Kit".
–
'ZV-Z21 typical'
denotes the cal kit "ZV-Z21 typical".
VNA command reference