VNA GUI reference
R&S
®
ZNL/ZNLE
502
User Manual 1178.5966.02 ─ 19
Standards with unrestricted port assignment ("any") are stored with their gender.
When a connector type and calibration kit are selected for the calibration, the ana-
lyzer checks whether the kit contains the required standard types and whether the
standards have the right gender.
Standards with restricted port assignment are always assumed to have the gender
that is appropriate for the calibrated port. The
port assignment
is stored in the cal-
ibration kit file, instead of the gender. During the calibration, the analyzer checks
whether the cal kit contains the necessary standard types for the required ports.
Remote command:
[SENSe:]CORRection:CKIT:<StandardType>
[SENSe:]CORRection:CKIT:<ConnType>:SELect
Add / Copy... / Delete / View / Modify...
The buttons in the right part of the dialog are used to manage standards:
●
"Add" adds a new standard to the calibration kit. The properties of the standard can
be edited in the table.
●
"Copy..." creates a standard based on the properties of an existing standard.
●
"Delete" deletes the selected standard.
●
"View / Modify..." opens the "View / Modify Cal Kit Standards" dialog. This dialog
shows the circuit model for the selected standard. For user-defined standard, you
can modify the circuit model parameters. See
"View / Modify Cal Kit Standards dia-
Remote command:
[SENSe:]CORRection:CKIT:<StandardType>
[SENSe:]CORRection:CKIT:<StandardType>:WLABel
Read .s1p File... / Read .s2p File... /
Opens a file selection dialog where you can select a Touchstone file containing the
reflection or transmission S-parameters for the standard. The R&S
ZNL/ZNLE uses the
imported S-parameters rather than the circuit model to characterize the standard, if
".s1p File"/".s2p File" is checked in the standard table. The appropriate file type (
*.s1p
for one-port standards and
*.s2p
for two-port standards) is selected automatically.
Remote command:
Write .s1p File/Write .s2p File
This button is only enabled, if the selected 1-port/2-port standard is defined using an
s1p/s2p file. It opens a dialog that allows you to export the respective s<n>p file.
This feature significantly facilitates the creation of custom cal kits.
Remote command:
n.a.
View / Modify Cal Kit Standards dialog
The "View / Modify Cal Kit Standards" dialog shows the circuit model for a selected cal-
ibration standard. It is also used to define or edit the circuit model (offset and load)
parameters for a user-defined standard.
Cal softtool