Command reference
R&S
®
ZNL/ZNLE
698
User Manual 1178.5966.02 ─ 19
Target Format
Description
Formula
SWR
(Voltage) Standing Wave Ratio
SWR = (1 + |z|) / (1 – |z|)
DEFault
Identical to trace format.
Note
: the Smith and Polar traces
use "Lin Mag" as the default for-
mat for target value.
-
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk>
Marker number (ignored for reference markers)
Parameters:
<SearchFormat>
MLINear | MLOGarithmic | PHASe | UPHase | REAL |
IMAGinary | SWR | DEFault
The format of the target value in the marker target search
(
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute TAR
CALCulate<Chn>:MARKer<Mk>:REFerence:
*RST:
DEFault
Example:
Suppose channel 1's selected trace is POLar and marker 1 isn't
yet created
:CALCULATE1:MARKER1 ON
Create/enable Marker 1
:CALCulate1:MARKer1:FUNCtion:SELect TARGet
Select TARGet search mode for marker 1
:CALCulate1:MARKer1:SEARch:FORMat?
Query the target format of marker 1. The result is DEF and for
polar diagrams the default target format is "Phase".
:CALCulate1:MARKer1:FUNCtion:TARGet?
Query for the default target value; for "Phase" this is 0 (degrees)
:CALCulate1:MARKer1:SEARch:FORMat MLOGarithmic
Change the target search format to dB magnitude
:CALCulate1:MARKer1:FUNCtion:TARGet?
Query for the default target value; for dB magnitude this is 0 (dB)
:CALCulate1:MARKer1:FUNCtion:TARGet -3
Set the target value to -3 dB
:CALCulate1:MARKer1:FUNCtion:EXECute
Execute the target search for marker 1
:CALCulate1:MARKer1:FUNCtion:RESult?
Query for the results.
Manual operation:
See
CALCulate<Chn>:MARKer<Mk>:SEARch:TRACking
<Boolean>
CALCulate<Chn>:MARKer<Mk>:REFerence:SEARch:TRACking
<Boolean>
These commands enable or disable tracking for the respective marker (see
VNA command reference