VNA GUI reference
R&S
®
ZNL/ZNLE
344
User Manual 1178.5966.02 ─ 19
initiates a new measurement cycle.
This setting applies to the active channel.
Remote command:
See also:
CONFigure:CHANnel<Ch>:MEASure[:STATe]
CONFigure:CHANnel:MEASure:ALL[:STATe]
Sweeps
Defines the number of sweeps to be performed in "Single" sweep mode (see
This setting applies to the active channel. In "Continuous" sweep mode, it is not availa-
ble (grayed out).
Remote command:
[SENSe<Ch>:]SWEep:COUNt
[SENSe:]SWEep:COUNt:ALL
Restart Sweep
Stops the current measurement cycle and starts a new one.
The effect of this command depends on the
of the active channel, and
the restart behavior defined in the
"Sweep Active Channel"
"Sweep All Channels"
Active channel in "Continuous"
sweep mode
The "Continuous" sweep mode
channels are swept cyclically, the
"Single" sweep channels are on
hold.
First the "Single" sweep mode
channels are swept one after the
other, each with its channel-spe-
cific number of "Sweeps". Then
the "Continuous" sweep mode
channels are sweep cyclically.
Active channel in "Single"
sweep mode
First the active channel is swept
with its channel-specific number
of "Sweeps". Then the "Continu-
ous" sweep mode channels are
sweep cyclically.
Remote command:
INITiate<Ch>[:IMMediate][:DUMMy]
INITiate[:IMMediate]:ALL
Restart Manager...
.
All Channels Continuous/All Channels on Hold
Selects a common
for all channels of the active channel setup.
●
"All Channels Continuous": all channels are set to "Continuous" sweep mode
●
"All Channels on Hold": all channels are set to "Single" sweep mode.
initiates a new measurement cycle.
Remote command:
Sweep Softtool