VNA GUI reference
R&S
®
ZNL/ZNLE
343
User Manual 1178.5966.02 ─ 19
For example, a new sweep can be triggered by an external trigger no. 1, while the indi-
vidual sweep points are triggered by external trigger no. 2. External trigger 1 is ignored
if a sweep is running, external trigger 2 is ignored if there is no running sweep. In the
figure below a sweep comprises 5 measurement points and dotted arrows depict
ignored trigger events.
Remote command:
TRIGger<Ch>[:SEQuence]:MULTiple:SOURce
TRIGger<Ch>[:SEQuence]:MULTiple:SLOPe<Num>
TRIGger<Ch>[:SEQuence]:MULTiple:HOLDoff
10.5.4
Sweep Control tab
Allows you to control the sweep execution.
10.5.4.1
Controls on the Sweep Control tab
Continuous / Single
Activate either continuous (default) or single sweep mode.
●
In "Continuous" sweep mode, the analyzer measures the related channel (or chan-
nels) continuously, repeating the current sweep over and over.
●
In "Single" sweep mode, the measurement is stopped after the configured number
of
Sweep Softtool