Command reference
R&S
®
ZNL/ZNLE
966
User Manual 1178.5966.02 ─ 19
........................................................................................... 969
..................................................................................970
...........................................................................................971
[SENSe<Ch>:]SWEep:COUNt
<Sweep>
Defines the channel-specific number of sweeps to be measured and buffered in single
sweep mode (
to define the sweep count for all channels.
Suffix:
<Ch>
.
Channel number
Parameters:
<Sweep>
Number of consecutive sweeps to be measured and buffered.
Range:
1 to 100000
*RST:
1
Example:
See
CALCulate<Chn>:DATA:NSWeep:FIRSt?
Manual operation:
See
[SENSe:]SWEep:COUNt:ALL
<Sweep>
Defines the number of sweeps to be measured in single sweep mode
(
OFF
). The setting is applied to all channels. Use
to define the sweep count for a single channel.
Setting parameters:
<Sweep>
Number of consecutive sweeps to be measured.
Range:
1 to 100000
*RST:
1
Example:
See
Chapter 11.6.2.4, "Data handling"
Usage:
Setting only
Manual operation:
See
[SENSe<Ch>:]SWEep:DWELl
<MeasDelay>
Defines the "Meas. Delay" before each partial measurement or the first partial mea-
surement (see
[SENSe<Ch>:]SWEep:DWELl:IPOint
). Setting a delay disables the
automatic calculation of the (minimum) sweep time (see
).
Suffix:
<Ch>
.
Channel number
VNA command reference