Command reference
R&S
®
ZNL/ZNLE
663
User Manual 1178.5966.02 ─ 19
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<TimeGateFilter>
BPASs | NOTCh
BPASs
Band pass filter: Pass all information in specified time region and
reject everything else
NOTCh
Notch filter: Reject all information in specified time region and
pass everything else
*RST:
BPASs
Example:
*RST; :CALC:FILT:TIME:STAT ON
Reset the instrument and enable the time gate.
CALC:FILT:TIME NOTCh
Select a notch filter in order to reject unwanted pulses.
Manual operation:
See
CALCulate<Chn>:FILTer[:GATE]:TIME:WINDow
<TimeGate>
Selects the time gate to be applied to the time domain transform.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<TimeGate>
RECT | HAMMing | HANNing | BOHMan | DCHebyshev
RECT - steepest edges (rectangle)
HANN - normal gate (Hann)
HAMMing - steep edges (Hamming)
BOHMan - minimum flatness (Bohman)
DCHebyshev - arbitrary gate shape (Dolph-Chebychev)
*RST:
HANN
Example:
See
CALCulate<Chn>:FILTer[:GATE]:TIME:DCHebyshev
Manual operation:
See
CALCulate:LIMit...
The
CALCulate:LIMit...
commands define the limit lines and control the limit
check.
CALCulate<Chn>:LIMit:CIRCle:CLEar
..............................................................................664
CALCulate<Chn>:LIMit:CIRCle:DATA
...............................................................................664
CALCulate<Chn>:LIMit:CIRCle:DISPlay[:STATe]
............................................................... 665
CALCulate<Chn>:LIMit:CIRCle:FAIL?
.............................................................................. 665
CALCulate:LIMit:CIRCle:FAIL:ALL?
................................................................................. 666
CALCulate<Chn>:LIMit:CIRCle:SOUNd[:STATe]
................................................................666
CALCulate<Chn>:LIMit:CIRCle[:STATe]
............................................................................666
VNA command reference