Command reference
R&S
®
ZNL/ZNLE
737
User Manual 1178.5966.02 ─ 19
Example:
CALCulate:TRANsform:DTFault:DEFine 'My cable
type', 1.1, 0.01, 1GHz, 0.015, 2GHz, 0.012,
3GHz
Defines a new cable type "My cable type" with relative permittiv-
ity ε
r
=1.1 and the frequency-dependent attenuation given in the
table below.
Usage:
Setting only
Options:
R&S
ZNL-K3
Manual operation:
See
Table 11-8: Example: frequency-dependent attenuation table
[i]
Frequency
Attenuation
1
1 GHz
0.01 dB/m
2
2 GHz
0.015 dB/m
3
3 GHz
0.012 dB/m
CALCulate<Chn>:TRANsform:DTFault:DELete
<DtfDeleteCable>
Deletes the user-defined cable type with the given name.
Suffix:
<Chn>
.
Channel number
This suffix is ignored: cable types are defined for (and deleted
from) all channels.
Setting parameters:
<DtfDeleteCable>
Name of a user-defined cable type.
Example:
CALCulate:TRANsform:DTFault:DELete 'My cable
type'
Deletes the user-defined cable type "My cable type".
Usage:
Setting only
Options:
R&S
ZNL-K3
Manual operation:
See
CALCulate<Chn>:TRANsform:DTFault:PEAK:COUNt?
If the active trace of channel <Chn> is a Distance to Fault (DtF) trace and DtF limit
checking is enabled, this query returns the number of DtF limit violations of this trace.
Use
CALCulate<Chn>:TRANsform:DTFault:PEAK:STATe
to enable DtF limit
checking and
CALCulate<Chn>:TRANsform:DTFault:PEAK:THReshold
the fault limit.
Suffix:
<Chn>
.
Channel number used to identify the active trace
VNA command reference