Remote Commands
R&S
®
FSW
982
User Manual 1173.9411.02 ─ 43
Usage:
Query only
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:MEAN:RESult?
This command queries the mean time domain power.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
.
<m>
Return values:
<MeanPower>
Mean power of the signal during the measurement time.
Usage:
Query only
Manual operation:
See
Measuring the Peak Power
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PPEak:AVERage:RESult?
This command queries the average positive peak time domain power. The query is
only possible if averaging has been activated previously using
MARKer<m>:FUNCtion:SUMMary:AVERage
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
.
<m>
Return values:
<PeakPower>
Peak power of the signal during the measurement time.
Usage:
Query only
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PPEak:PHOLd:RESult?
This command queries the maximum positive peak time domain power. The query is
only possible if the peak hold function has been activated previously using
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PHOLd
.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.