Common Measurement Settings
R&S
®
FSW
459
User Manual 1173.9411.02 ─ 43
5.
To detect a spurious signal close to the noise floor:
● Set the "RF Attenuation" to "Manual" mode and reduce the "Value" to lower the
noise floor.
● Select "Relative" - "Logarithmic" scaling.
Now you can determine if any spurious levels of a certain size are visible.
8.5
Bandwidth, Filter and Sweep Configuration
The basic bandwidth, filter and sweep settings that apply to most measurements are
described here. These parameters define how the data is measured: how much data is
collected internally and which filters are used.
Impact of the Bandwidth, Filter and Sweep Settings
.............................................459
Bandwidth, Filter and Sweep Settings
.................................................................. 465
Reference: List of Available RRC and Channel Filters
......................................... 474
8.5.1
Impact of the Bandwidth, Filter and Sweep Settings
The bandwidth, filter and sweep settings are closely related and interdependent. The
values available for resolution bandwidth and video bandwidth depend on the selected
filter type. In addition, these settings have an impact on other measurement parame-
ters. The following equation shows the interdependency of these settings:
T
MIN
= K*Span/RBW
2
where K = Filter constant
By default, a Gaussian filter is used. The resolution bandwidth, the video bandwidth
and the "Sweep Time" are set automatically according to the set span, and default cou-
pling is used. Thus, the following settings are applied:
RBW = 100 * Span
VBW = RBW = 100 * Span
"Sweep Time" = T
min
for set Span, RBW, VBW
When defining the bandwidth and filter settings, consider the impact of the individual
settings on the other settings and the measurement result, as described in more detail
in the following sections.
Separating Signals by Selecting an Appropriate Resolution Bandwidth
Smoothing the Trace Using the Video Bandwidth
.................................................460
...................................................................................... 461
...................................................................................... 461
How Data is Measured: the Sweep Type
..............................................................462
Which Data May Pass: Filter Types
...................................................................... 463
How Long the Data is Measured: Sweep Time
.....................................................464
How Much Data is Measured: Sweep Points and Sweep Count
.......................... 464
How Often Data is Measured: Sweep Mode
.........................................................465
Bandwidth, Filter and Sweep Configuration
Содержание FSW Series
Страница 1: ...R S FSW Signal and Spectrum Analyzer User Manual User Manual Version 43 1173941102 2...
Страница 16: ...Contents R S FSW 16 User Manual 1173 9411 02 43...
Страница 1388: ...Maintenance R S FSW 1388 User Manual 1173 9411 02 43 c If necessary remove any excess moisture with a dry soft cloth Cleaning...