Getting Started
R&S
®
FSW
67
User Manual 1173.9411.02 ─ 43
Figure 5-7: Spectrogram of the calibration signal
5.3.3
Activating Additional Measurement Channels
The R&S
FSW features multiple measurement channels, i.e. you can define several
measurement configurations in parallel and then switch between the channels auto-
matically to perform the measurements sequentially. We will demonstrate this feature
by activating additional measurement channels for a different frequency range, a zero
span measurement, and an I/Q analysis.
To activate additional measurement channels
1. Press the [Mode] key on the front panel.
2. On the "New Channel" tab of the " Spectrum Mode" dialog box, tap the
"Spectrum" button.
Trying Out the Instrument
Содержание FSW Series
Страница 1: ...R S FSW Signal and Spectrum Analyzer User Manual User Manual Version 43 1173941102 2...
Страница 16: ...Contents R S FSW 16 User Manual 1173 9411 02 43...
Страница 1388: ...Maintenance R S FSW 1388 User Manual 1173 9411 02 43 c If necessary remove any excess moisture with a dry soft cloth Cleaning...