Remote Commands
R&S
®
FSW
1156
User Manual 1173.9411.02 ─ 43
Example:
CALC:MARK3 ON
Switches on marker 3.
Manual operation:
See
See
See
CALCulate<n>:MARKer<m>:TRACe
<Trace>
This command selects the trace the marker is positioned on.
Note that the corresponding trace must have a trace mode other than "Blank".
If necessary, the command activates the marker first.
Suffix:
<n>
.
<m>
Parameters:
<Trace>
1 to 6
Trace number the marker is assigned to.
Example:
//Assign marker to trace 1
CALC:MARK3:TRAC 2
Manual operation:
See
"Assigning the Marker to a Trace"
CALCulate<n>:MARKer<m>:X
<Position>
This command moves a marker to a specific coordinate on the x-axis.
If necessary, the command activates the marker.
If the marker has been used as a delta marker, the command turns it into a normal
marker.
Suffix:
<n>
.
<m>
Parameters:
<Position>
Numeric value that defines the marker position on the x-axis.
The unit depends on the result display.
Range:
The range depends on the current x-axis range.
Default unit: Hz
Example:
CALC:MARK2:X 1.7MHz
Positions marker 2 to frequency 1.7 MHz.
Manual operation:
See
"Marker 1/Marker 2/Marker 3/Marker 4"
See
See
See
See
Analyzing Measurements (Basics)
Содержание FSW Series
Страница 1: ...R S FSW Signal and Spectrum Analyzer User Manual User Manual Version 43 1173941102 2...
Страница 16: ...Contents R S FSW 16 User Manual 1173 9411 02 43...
Страница 1388: ...Maintenance R S FSW 1388 User Manual 1173 9411 02 43 c If necessary remove any excess moisture with a dry soft cloth Cleaning...