Remote Commands
R&S
®
FSW
915
User Manual 1173.9411.02 ─ 43
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
0
Example:
CONF:GEN:RFO:STAT ON
Manual operation:
See
CONFigure:GENerator:TARGet:PATH:BB?
Queries the BB signal path of the generator used for signal generation.
Return values:
<Path>
A | B
Example:
CONF:GEN:TARG:PATH:BB?
Usage:
Query only
Manual operation:
See
CONFigure:GENerator:TARGet:PATH:RF
<Path>
Selects the RF signal path of the generator to be used for signal generation.
Parameters:
<Path>
A | B
Example:
CONF:GEN:TARG:PATH:RF B
Manual operation:
See
CONFigure:SETTings:NPRatio
Usage:
Event
Manual operation:
See
"Query all Generator Setup Settings from Generator"
CONFigure:SETTings:NPRatio:NOTCh
Usage:
Event
Manual operation:
See
"Query all Notch Settings from Generator"
14.5.6.4
Retrieving Results from an NP Measurement
..................................................................................... 915
CALCulate<n>:NPRatio:RESult?
<ResultType>
Queries the power results of the noise power ratio measurement.
Configuring and Performing Measurements