Remote Commands
R&S
®
FSW
1112
User Manual 1173.9411.02 ─ 43
The power sensor uses the duty cycle in combination with the mean power to calculate
the power of the pulse.
Suffix:
<p>
.
Power sensor
Parameters:
<Percentage>
Range:
0.001 to 99.999
*RST:
99.999
Default unit: %
Example:
PMET2:DCYC:STAT ON
Activates the duty cycle correction.
PMET2:DCYC:VAL 0.5
Sets the correction value to 0.5%.
Manual operation:
See
[SENSe:]PMETer<p>:FREQuency
<Frequency>
This command defines the frequency of the power sensor.
Suffix:
<p>
.
Power sensor index
Parameters:
<Frequency>
The available value range is specified in the data sheet of the
power sensor in use.
*RST:
50 MHz
Default unit: HZ
Example:
PMET2:FREQ 1GHZ
Sets the frequency of the power sensor to 1 GHz.
Manual operation:
See
[SENSe:]PMETer<p>:FREQuency:LINK
<Coupling>
This command selects the frequency coupling for power sensor measurements.
Suffix:
<p>
.
Power sensor index
Parameters:
<Coupling>
CENTer
Couples the frequency to the center frequency of the analyzer
MARKer1
Couples the frequency to the position of marker 1
OFF
Switches the frequency coupling off
*RST:
CENTer
Example:
PMET2:FREQ:LINK CENT
Couples the frequency to the center frequency of the analyzer
Setting Basic Measurement Parameters
Содержание FSW Series
Страница 1: ...R S FSW Signal and Spectrum Analyzer User Manual User Manual Version 43 1173941102 2...
Страница 16: ...Contents R S FSW 16 User Manual 1173 9411 02 43...
Страница 1388: ...Maintenance R S FSW 1388 User Manual 1173 9411 02 43 c If necessary remove any excess moisture with a dry soft cloth Cleaning...