Remote Commands
R&S
®
FSW
1069
User Manual 1173.9411.02 ─ 43
AIQI
Analog Baseband I connector
This setting is only available if the Analog Baseband interface
(R&S
FSW-B71) is installed and active for input. It is not availa-
ble for the R&S
FSW67 or R&S
FSW85.
RFPRobe
Active RF probe
*RST:
RF
Example:
INP:CONN:RF
Selects input from the RF input connector.
Manual operation:
See
INPut<ip>:COUPling
<CouplingType>
This command selects the coupling type of the RF input.
Suffix:
<ip>
.
1 | 2
For R&S
FSW85 models with two RF input connectors:
1: Input 1 (1
mm [RF Input] connector)
2: Input 2 (1.85
mm [RF2 Input] connector)
For all other models:
irrelevant
Parameters:
<CouplingType>
AC | DC
AC
AC coupling
DC
DC coupling
*RST:
AC
Example:
INP:COUP DC
Manual operation:
See
INPut<ip>:DPATh
<State>
Enables or disables the use of the direct path for frequencies close to 0
Hz.
Suffix:
<ip>
.
1 | 2
For R&S
FSW85 models with two RF input connectors:
1: Input 1 (1
mm [RF Input] connector)
2: Input 2 (1.85
mm [RF2 Input] connector)
For all other models:
irrelevant
Parameters:
<State>
AUTO | OFF
Setting Basic Measurement Parameters
Содержание FSW Series
Страница 1: ...R S FSW Signal and Spectrum Analyzer User Manual User Manual Version 43 1173941102 2...
Страница 16: ...Contents R S FSW 16 User Manual 1173 9411 02 43...
Страница 1388: ...Maintenance R S FSW 1388 User Manual 1173 9411 02 43 c If necessary remove any excess moisture with a dry soft cloth Cleaning...