Remote Commands
R&S
®
FSW
1272
User Manual 1173.9411.02 ─ 43
For details, see
Chapter 12.4, "Frequency Response Correction (R&S
Remote Commands for Frequency Response Correction
.................................. 1272
Programming Example: Using Touchstone files
..................................................1288
14.10.4.1
Remote Commands for Frequency Response Correction
The following commands are only available if the Frequency Response Correction
option (R&S
FSW-K544) is installed.
Input-specific frequency correction
Frequency response correction can be configured for all inputs, or for particular input
types only. Be sure to use the correct command for the required setting.
●
[SENSe:]CORRection:FRESponse:INPut<ip>:USER:
commands are applied
to RF input only
●
[SENSe:]CORRection:FRESponse:BASeband:USER:
commands are applied
to baseband input only
●
[SENSe:]CORRection:FRESponse:USER:
commands are applied to all input
types; queries refer to the currently active input source
[SENSe:]CORRection:FRESponse<si>:BASeband:USER:FLISt<fli>:CATalog?
[SENSe:]CORRection:FRESponse<si>:INPut<ip>:USER:FLISt<fli>:CATalog?
[SENSe:]CORRection:FRESponse<si>:USER:FLISt<fli>:CATalog?
...................................1274
[SENSe:]CORRection:FRESponse<si>:BASeband:USER:FLISt<fli>:CLEar
[SENSe:]CORRection:FRESponse<si>:INPut<ip>:USER:FLISt<fli>:CLEar
[SENSe:]CORRection:FRESponse<si>:USER:FLISt<fli>:CLEar
........................................1274
[SENSe:]CORRection:FRESponse<si>:USER:FLISt<fli>:DATA:FREQuency?
[SENSe:]CORRection:FRESponse<si>:USER:FLISt<fli>:DATA:MAGNitude?
[SENSe:]CORRection:FRESponse<si>:USER:FLISt<fli>:DATA:PHASe?
........................... 1274
[SENSe:]CORRection:FRESponse<si>:BASeband:USER:FLISt<fli>:INSert
[SENSe:]CORRection:FRESponse<si>:INPut<ip>:USER:FLISt<fli>:INSert
[SENSe:]CORRection:FRESponse<si>:USER:FLISt<fli>:INSert
........................................1275
[SENSe:]CORRection:FRESponse<si>:BASeband:USER:FLISt<fli>:MAGNitude[:STATe]
[SENSe:]CORRection:FRESponse<si>:INPut<ip>:USER:FLISt<fli>:MAGNitude[:STATe]
[SENSe:]CORRection:FRESponse<si>:USER:FLISt<fli>:MAGNitude[:STATe]
[SENSe:]CORRection:FRESponse<si>:BASeband:USER:FLISt<fli>:PHASe[:STATe]
[SENSe:]CORRection:FRESponse<si>:INPut<ip>:USER:FLISt<fli>:PHASe[:STATe]
[SENSe:]CORRection:FRESponse<si>:USER:FLISt<fli>:PHASe[:STATe]
.......................... 1276
[SENSe:]CORRection:FRESponse<si>:BASeband:USER:FLISt<fi>:REMove
[SENSe:]CORRection:FRESponse<si>:INPut<ip>:USER:FLISt<fli>:REMove
[SENSe:]CORRection:FRESponse<si>:USER:FLISt<fli>:REMove
.................................... 1277
[SENSe:]CORRection:FRESponse<si>:BASeband:USER:FLISt<fli>:SELect
[SENSe:]CORRection:FRESponse<si>:INPut<ip>:USER:FLISt<fli>:SELect
[SENSe:]CORRection:FRESponse<si>:USER:FLISt<fli>:SELect
...................................... 1277
[SENSe:]CORRection:FRESponse<si>:BASeband:USER:FLISt<fli>:SIZE?
[SENSe:]CORRection:FRESponse<si>:INPut<ip>:USER:FLISt<fli>:SIZE?
[SENSe:]CORRection:FRESponse<si>:USER:FLISt<fli>:SIZE?
........................................1277
Configuring the R&S
FSW
Содержание FSW Series
Страница 1: ...R S FSW Signal and Spectrum Analyzer User Manual User Manual Version 43 1173941102 2...
Страница 16: ...Contents R S FSW 16 User Manual 1173 9411 02 43...
Страница 1388: ...Maintenance R S FSW 1388 User Manual 1173 9411 02 43 c If necessary remove any excess moisture with a dry soft cloth Cleaning...