Troubleshooting
R&S
®
FSW
1395
User Manual 1173.9411.02 ─ 43
The same applies to limit line settings.
If you want to apply the changed transducer values after recalling the save set you
must force the application to reload the transducer file. To do so, simply open the "Edit
Transducer" dialog box (see
Chapter 12.3.2, "Transducer Settings"
toggle the "X-Axis" option from "lin" to "log" and back. Due to that change, the trans-
ducer file is automatically reloaded, and the changed transducer values are applied to
the current measurement. Now you can create a new save set with the updated trans-
ducer values.
Similarly, if you want to apply the changed limit values after recalling the save set you
must force the application to reload the limit file. To do so, simply open the "Edit Limit
Line" dialog box (see
Chapter 9.4.2.2, "Limit Line Settings and Functions"
on page 563) and toggle the "Y-Axis" unit. Due to that change, the limit line file is auto-
matically reloaded, and the changed limit values are applied to the current measure-
ment. Now a new save set with the updated limit values can be created.
16.5
System Recovery
For instruments running Windows 10
, the system drive is delivered with a recovery
partition that allows you to restore the original operating system image and firmware.
To restore the original operating system image and firmware
1. Open the Windows control panel.
2. Select "Update & Security" > "Recovery" > "Restart Now".
The "R&S Recovery Environment" starts.
3. In the "R&S Recovery Environment", select "Factory Default Restore".
The default image is restored.
4. Reboot the instrument.
After the default image is restored, upgrade to the desired firmware version (see
ter 12.6.4, "Firmware Updates"
16.6
Collecting Information for Support
If problems occur, the instrument generates error messages which in most cases will
be sufficient for you to detect the cause of an error and find a remedy.
Error messages are described in
Chapter 5.4.1.6, "Error Information"
In addition, our customer support centers are there to assist you in solving any prob-
lems that you may encounter with your R&S
FSW. We will find solutions more quickly
and efficiently if you provide us with the information listed below.
●
Windows Event Log Files
Collecting Information for Support