Remote Commands
R&S
®
FSW
1109
User Manual 1173.9411.02 ─ 43
Suffix:
<p>
.
Power sensor index
Parameters:
<Placeholder>
Currently not used
<Type>
Detected power sensor type, e.g. "NRP-Z81".
<Interface>
Interface the power sensor is connected to; always "USB"
<SerialNo>
Serial number of the power sensor assigned to the specified
index
Example:
SYST:COMM:RDEV:PMET2:DEF '','NRP-Z81','',
'123456'
Assigns the power sensor with the serial number '123456' to the
configuration "Power Sensor 2".
SYST:COMM:RDEV:PMET2:DEF?
Queries the sensor assigned to "Power Sensor 2".
Result:
'','NRP-Z81','USB','123456'
The NRP-Z81 power sensor with the serial number '123456' is
assigned to the "Power Sensor 2".
Manual operation:
See
Configuring Power Sensor Measurements
CALibration:PMETer<p>:ZERO:AUTO ONCE
..................................................................1109
CALCulate<n>:PMETer<p>:RELative[:MAGNitude]
.......................................................... 1110
CALCulate<n>:PMETer<p>:RELative[:MAGNitude]:AUTO ONCE
...................................... 1110
CALCulate<n>:PMETer<p>:RELative:STATe
................................................................... 1110
[SENSe:]PMETer<p>:DCYCle[:STATe]
............................................................................ 1111
[SENSe:]PMETer<p>:DCYCle:VALue
..............................................................................1111
..................................................................................1112
[SENSe:]PMETer<p>:FREQuency:LINK
..........................................................................1112
......................................................................................... 1113
[SENSe:]PMETer<p>:MTIMe:AVERage:COUNt
............................................................... 1113
[SENSe:]PMETer<p>:MTIMe:AVERage[:STATe]
...............................................................1113
[SENSe:]PMETer<p>:ROFFset[:STATe]
.......................................................................... 1114
........................................................................................ 1114
[SENSe:]PMETer<p>:UPDate[:STATe]
............................................................................ 1114
........................................................................................ 1115
...............................................................................1115
CALibration:PMETer<p>:ZERO:AUTO ONCE
This command zeroes the power sensor.
Note that you have to disconnect the signals from the power sensor input before you
start to zero the power sensor. Otherwise, results are invalid.
Setting Basic Measurement Parameters