Remote Commands
R&S
®
FSW
967
User Manual 1173.9411.02 ─ 43
CALC:PSE:SUBR 10
//Sets 10 peaks per range to be stored in the list.
//--------------Performing the Measurement-----
INIT:SPUR; *WAI
//Performs a spurious emission measurement and waits until the sweep has finished.
//---------------Retrieving Results-------------
CALC:LIM1:FAIL?
//Queries the result of the check for limit line 1.
TRAC? SPUR
//Queries the peak list of the spurious emission measurement.
14.5.9
Analyzing Statistics (APD, CCDF)
All remote control commands specific to statistical measurements are described here.
Activating Statistical Measurements
..................................................................... 967
Configuring Statistical Measurements
...................................................................968
Using Gate Ranges for Statistical Measurements
................................................ 969
..............................................................................................971
Performing a Statistical Measurement
.................................................................. 974
.................................................................................................974
Programming Example: Measuring Statistics
....................................................... 975
14.5.9.1
Activating Statistical Measurements
The following commands activate statistical measurements.
CALCulate<n>:STATistics:APD[:STATe]
............................................................................967
CALCulate<n>:STATistics:CCDF[:STATe]
......................................................................... 967
CALCulate<n>:STATistics:APD[:STATe]
<State>
This command turns the APD measurement on and off.
Suffix:
<n>
.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
CALC:STAT:APD ON
Switches on the APD measurement.
CALCulate<n>:STATistics:CCDF[:STATe]
<State>
This command turns the CCDF on and off.
Configuring and Performing Measurements