Remote Commands
R&S
®
FSW
969
User Manual 1173.9411.02 ─ 43
Parameters:
<Samples>
Range:
Min: 100, Max: depends on the RBW filter
*RST:
100000
Example:
CALC:STAT:NSAM 500
Sets the number of measurement points to be acquired to 500.
Manual operation:
See
14.5.9.3
Using Gate Ranges for Statistical Measurements
The following commands control gated statistical measurements.
[SENSe:]SWEep:EGATe:TRACe<t>:COMMent
................................................................. 969
[SENSe:]SWEep:EGATe:TRACe<t>:PERiod
.....................................................................969
[SENSe:]SWEep:EGATe:TRACe<t>:STARt<gr>
................................................................ 970
[SENSe:]SWEep:EGATe:TRACe<t>[:STATe<gr>]
.............................................................. 970
[SENSe:]SWEep:EGATe:TRACe<t>:STOP<gr>
................................................................ 970
[SENSe:]SWEep:EGATe:TRACe<t>:COMMent
<Comment>
This command defines a comment for the gate of a particular trace.
Suffix:
<t>
.
Parameters:
<Comment>
String containing the comment.
Example:
SWE:EGAT:TRAC1:COMM 'MyComment'
Defines a comment for the gate in trace 1.
Manual operation:
See
[SENSe:]SWEep:EGATe:TRACe<t>:PERiod
<Length>
This command defines the length of the gate for all traces.
The gate length applies to all traces.
Suffix:
<t>
.
irrelevant
Parameters:
<Length>
Range:
100 ns to 1000 s
*RST:
2 ms
Default unit: s
Example:
SWE:EGAT:TRAC:PER 5ms
Defines the period for gated triggering to 5 ms.
Manual operation:
See
Configuring and Performing Measurements