Remote Commands
R&S
®
FSW
875
User Manual 1173.9411.02 ─ 43
Example:
See
"Example: Configuring and Performing an Asymmetrical
CALCulate<n>:LIMit<li>:ACPower:GAP<gap>:MANual:LOWer:ACLR[:RELative]
<SBGaps>, <Limit>
CALCulate<n>:LIMit<li>:ACPower:GAP<gap>:MANual:LOWer:ACLR[:RELative]?
<SBGaps>
This command defines the relative limit for the ACLR power in the specified lower gap
channel. The reference value for the relative limit is the measured channel power.
If you define both an absolute limit and a relative limit, the R&S
FSW uses the lower
value for the limit check.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
<gap>
1 | 2
Gap channel number
Parameters:
<Limit>
Defines the relative limit for the ACLR power in the specified gap
channel.
Default unit: DB
Parameters for setting and query:
<SBGaps>
AB | BC | CD | DE | EF | FG | GH
Name of the gap, defined by the letters of the surrounding sub
blocks (e.g. "AB" for the gap between sub blocks A and B).
Example:
CALC:LIM:ACP:GAP1:MAN:LOW:ACLR:REL AB,3dB
Example:
See
"Example: Configuring and Performing an Asymmetrical
CALCulate<n>:LIMit<li>:ACPower:GAP<gap>:MANual:LOWer:ACLR[:RELative]:
STATe
<SBGaps>, <State>
CALCulate<n>:LIMit<li>:ACPower:GAP<gap>:MANual:LOWer:ACLR[:RELative]:
STATe?
<SBGaps>
This command turns the relative limit check for the specified lower gap (CACLR) chan-
nel on and off.
You have to activate the general ACLR limit check before using this command with
CALCulate<n>:LIMit<li>:ACPower[:STATe]
.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
Configuring and Performing Measurements
Содержание FSW Series
Страница 1: ...R S FSW Signal and Spectrum Analyzer User Manual User Manual Version 43 1173941102 2...
Страница 16: ...Contents R S FSW 16 User Manual 1173 9411 02 43...
Страница 1388: ...Maintenance R S FSW 1388 User Manual 1173 9411 02 43 c If necessary remove any excess moisture with a dry soft cloth Cleaning...