Remote Commands
R&S
®
FSW
850
User Manual 1173.9411.02 ─ 43
[SENSe:]POWer:ACHannel:FILTer[:STATe]:ALTernate<ch>
<State>
This command turns the weighting filter for an alternate channel on and off.
For asymmetrical MSR signals, this command turns the weighting filter for the lower
alternate channels on and off. To configure the filter state for the upper alternate chan-
nels, use the
[SENSe:]POWer:ACHannel:FILTer[:STATe]:UALTernate<ch>
command.
Suffix:
<ch>
.
1..n
Alternate channel number
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Manual operation:
See
See
[SENSe:]POWer:ACHannel:FILTer[:STATe]:CHANnel<ch>
<State>
This command turns the weighting filter for a transmission channel on and off.
Suffix:
<ch>
.
1..n
Tx channel number
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Manual operation:
See
14.5.3.4
Selecting the Reference Channel
The following commands define the reference channel for relative ACLR measure-
ments.
[SENSe:]POWer:ACHannel:REFerence:AUTO ONCE
....................................................... 850
[SENSe:]POWer:ACHannel:REFerence:TXCHannel:AUTO
................................................ 851
[SENSe:]POWer:ACHannel:REFerence:TXCHannel:MANual
............................................. 851
[SENSe:]POWer:ACHannel:REFerence:AUTO ONCE
This command sets the channel power as the reference for relative ACLR measure-
ments.
Example:
POW:ACH:REF:AUTO ONCE
Usage:
Event
Manual operation:
See
"Setting a fixed reference for Channel Power measure-
Configuring and Performing Measurements