Common Analysis and Display Functions
R&S
®
FSW
563
User Manual 1173.9411.02 ─ 43
Storing and Recalling Limit Lines
Limit lines can be stored with the configuration settings so they can be recalled for
other measurements at a later time (see
Chapter 11.3, "Storing and Recalling Instru-
ment Settings and Measurement Data"
on page 625). Note, however, that any
changes made to the limit lines
after
storing the configuration file cannot be restored
and will be overwritten by the stored values when the configuration file is recalled.
Always remember to store the settings again after changing the limit line values.
After recalling measurement settings, the limit line values applied to the measurement
may be different to those displayed in the "Limit Lines" dialog box; see
recalling transducer and limit line settings"
9.4.2.2
Limit Line Settings and Functions
Access
: "Overview" > "Analysis" > "Lines"
or
: [LINES] > "Line Config"
Up to 8 limit lines can be displayed simultaneously in the R&S
FSW. Many more can
be stored on the instrument.
Stored limit line settings
When storing and recalling limit line settings, consider the information provided in
ing and recalling transducer and limit line settings"
.........................................................................................563
Limit Line Management
Access
: "Overview" > "Analysis" > "Lines" > "Limit Lines"
or
: [LINES] > "Line Config" > "Limit Lines"
Display and Limit Lines