Measurements and Results
R&S
®
FSW
333
User Manual 1173.9411.02 ─ 43
Initial peak search results - Marker Table
As a result of the initial peak search, the active markers are set to the positive peaks of
the measured signal.
If
Auto peak search and limit lines
are active, the active markers are set to the peak
delta
values between the measured signal and the limit lines.
The active marker levels and positions are displayed in the
Marker Table
.
(Note: the marker results are also displayed in the Result Summary; in addition, the
Marker Table contains the marker results for those markers for which no final EMI test
is performed.)
Final test results - Result Summary
The results of the final EMI tests at the active marker frequencies are displayed in the
Result Summary
.
The Result Summary provides the following information:
Label
Description
Type
Marker name
Ref
Reference marker for delta markers
Trace
Assigned trace
X-value
Marker x-value (frequency for final test)
Y-value
Marker y-value (level during inital measurement)
Final Test
Detector used for final EMI test
Line name
Line activated for limit check
Δ
Limit
Delta between measured level and limit line (if active)
The value is colored to indicate the following states:
●
green
: does not exceed limit
●
yellow
: within margin
●
red
: exceeds limit
Final Result
Value measured during final EMI test using specified detector at marker frequency
7.13.3
EMI Measurement Basics
Some background knowledge on basic terms and principles used in EMI measure-
ments is provided here for a better understanding of the required configuration set-
tings.
Resolution Bandwidth and Filter Types
.................................................................334
..................................................................................... 334
Frequency Resolution - Sweep Points and Scaling
.............................................. 338
.............................................................................. 339
.....................................................................................339
Electromagnetic Interference (EMI) Measurement