Measurements and Results
R&S
®
FSW
249
User Manual 1173.9411.02 ─ 43
Table 7-13: Limit lines in overlapping ranges crossing the mid-frequency
Initial situation
: overlapping ranges
Range 4 ("None") + Range 5 ("None") overlap and cross the
mid-frequency between sub blocks 1 and 2
Result:
Subranges 4a and 5a are created left and right of the
mid-frequency;
For subrange 4a: limit line and parameters of range 4 apply
For subrange 5a: limit line and parameters of range 5 apply
Global SEM limit check
For the complete SEM measurement, which can consist of multiple sub blocks, only
one single limit check is performed. A single limit line is calculated according to the
individual range limit lines and the defined functions for overlapping ranges. The mea-
sured values are then compared with this single limit line. If the limit is exceeded in any
range, the result of the limit check is ."' failed!"