Common Measurement Settings
R&S
®
FSW
493
User Manual 1173.9411.02 ─ 43
"Level"
The trigger event for the gate to open is a particular power level.
After the gate signal has been detected, the gate remains open until
the signal disappears.
Note:
If you perform gated measurements in combination with the IF
Power trigger, the R&S
FSW ignores the holding time for frequency
sweep, FFT sweep, zero span and I/Q mode measurements.
This mode is not supported when using R&S Power Sensors as
power triggers ("Trg/Gate Source" =
Power Sensor
or
External
).
Remote command:
Gate Delay
Defines the delay time between the gate signal and the continuation of the measure-
ment.
The delay position on the time axis in relation to the sweep is indicated by a line
labeled "GD".
For more information see
Chapter 8.6.2.1, "Gated Measurements"
Remote command:
Gate Length
Defines how long the gate is open when it is triggered.
The gate length can only be set in the edge-triggered gate mode. In the level-triggered
mode the gate length depends on the level of the gate signal.
The gate length in relation to the sweep is indicated by a line labeled "GL".
For more information see
Chapter 8.6.2.1, "Gated Measurements"
Remote command:
8.6.2.3
Continuous Gate Settings
Access
: "Overview" > "Trigger" > "Trigger / Gate Config." > "Cont. Gate" tab
Continuous gating allows you to perform a continuous gated sweep after a single
external trigger is received.
Trigger and Gate Configuration
Содержание FSW Series
Страница 1: ...R S FSW Signal and Spectrum Analyzer User Manual User Manual Version 43 1173941102 2...
Страница 16: ...Contents R S FSW 16 User Manual 1173 9411 02 43...
Страница 1388: ...Maintenance R S FSW 1388 User Manual 1173 9411 02 43 c If necessary remove any excess moisture with a dry soft cloth Cleaning...