Remote Commands
R&S
®
FSW
953
User Manual 1173.9411.02 ─ 43
//------------ Configuring List Evaluation-----------------
//Activate list evaluation, i.e. the peak is determined for each range
//after each sweep.
CALC:ESP:PSE:AUTO ON
//Define a peak threshold of 10 dB.
CALC:ESP:PSE:MARG 10dB
//------------ Managing Measurement Configurations----------
//Save the current configuration in a new file named '3GPP_UL_User'
//in the same directory so the standard is not overwritten.
ESP:PRES:STOR 'WCDMA\3GPP\UL\3GPP_UL_User.xml'
//------------ Performing the measurement-------------------
//One sweep
INIT:ESP
//------------ Checking the Results-------------------------
//Query the result of the limit check for all ranges.
CALC:LIM:FAIL?
//Query the peak for each range of the SEM measurement as a list.
TRAC:DATA? LIST
14.5.8
Measuring Spurious Emissions
The following commands are required to perform spurious emissions measurements.
................................................................................. 953
...................................................................................... 954
Configuring the List Evaluation
............................................................................. 961
Adjusting the X-Axis to the Range Definitions
.......................................................963
Performing a Spurious Measurement
................................................................... 964
Retrieving and Saving Settings and Results
.........................................................964
Programming Example: Spurious Emissions Measurement
.................................964
14.5.8.1
Initializing the Measurement
Note that with the R&S
FSW, the spurious measurement must be initialized before you
can start configuring the sweep list or list evaluation.
INITiate<n>:SPURious
This command initiates a Spurious Emission measurement.
Suffix:
<n>
.
Configuring and Performing Measurements