Remote Commands
R&S
®
FSW
928
User Manual 1173.9411.02 ─ 43
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:ABSolute:STOP
<Level>
This command defines an absolute limit for a SEM range.
Unlike manual operation, you can define an absolute limit anytime and regardless of
the limit check mode.
Suffix:
<sb>
.
1 to 8
Sub block in a Multi-SEM measurement
<ri>
1..n
Selects the measurement range.
<li>
1..n
Power class for which the limit is defined.
Parameters:
<Level>
Absolute limit at the stop frequency of a SEM range.
Range:
-400 to 400
*RST:
-13
Default unit: dBm
Example:
SENSe:ESPectrum:RANGe:LIMit:ABSolute:STOP -15
Chapter 14.5.7.11, "Example: SEM
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:STARt
<Level>
This command defines a relative limit for a SEM range.
Unlike manual operation, you can define a relative limit regardless of the limit check
mode.
Suffix:
<sb>
.
1 to 8
Sub block in a Multi-SEM measurement
<ri>
1..n
Selects the SEM range.
<li>
1..n
Power class for which the limit is defined.
Parameters:
<Level>
Relative limit at the start frequency of a SEM range.
Range:
-400 to 400
*RST:
-50
Default unit: dBc
Example:
SENS:ESP:RANG:LIM:REL:STAR -10
Manual operation:
See
Configuring and Performing Measurements